HP-1100 injection program for performing trap/wash/elute

The injection program shown below is used with the HP-1100 to perform a clean/trap/wash sequence.  The program first performs 2 injections of  "protein Drano" (50% formic acid/50 % ACN or propanol) from vial #1, before injecting the defined amount from the sample vial.  The Drano step is used to minimize carryover between protein samples.

  1. VALVE bypass
  2. EJECT all into seat
  3. REPEAT 2 times
  4. DRAW 50uL from vial 1
  5. INJECT
  6. WAIT 0.5 min
  7. END REPEAT
  8. VALVE bypass
  9. EJECT all into seat
  10. DRAW def. amount from sample
  11. INJECT

---end of program---